Baker Hughes 旗下的 Panametrics 采用可调谐二极管激光器吸收光谱技术 (TDLAS) 来测量常见气体中的微量水含量。 Panametrics TDLAS 组合属于 Aurora 产品系列,内含配有直观界面和专利技术的工业分析仪,具备卓越性能。
这类可调谐二极管激光器分析仪因反应快速准确而闻名,能够分析 2-5000 PPMv 范围的水含量。凭借获得专利的校准方案,即使背景气体发生变化,此分析仪也能做到准确无误,而其他分析仪在这种情况下通常会发生错误。
我们的微量水分析仪配备了 TDLAS,无需定期重新校准,可降低客户的拥有成本。
Aurora 拥有以下三种配置:
- Aurora 能够对工艺异常做出快速反应并在恢复时快速干燥。
- Aurora TransPort 将 TDLAS 的卓越性能融入坚固便携的分析仪中,并且可以使用电池或在线电源。
- Aurora RM 机架安装式分析仪采用 TDLAS 来准确测量各种背景气体中的湿度。
Our industrial TDLAS analyzer, Aurora, provides moisture measurements from 0 to 5000 PPM. Its accurate measurement technique measures moisture in many applications such as natural gas TEG dehydration, natural gas storage, transmission and custody transfer, carbon dioxide for enhanced oil recovery and carbon sequestration, metal heat-treating furnace gases, and hydrogen recycle gas in reforming. Features include:
Range
- 0 - 5000 ppm measurement range, for CO2 applications: 0 - 1000 ppm
- Dew/Frost Point: -97º to 27.3ºF (-71º to -2.6ºC) frost point @ STP of 25ºC, 14.696 psia
- Absolute Humidity: 0.095 to 237 lbs/MMSFC (1.52 to 3,803 mg/m3)
Accuracy
- ±1% of reading or ±2 ppm, whichever is greater; for > 1000 ppm, ±5% of reading
- For CO2 applications: ±3% of reading or ±5 ppmv
- For H2 recycle applications: ±1% of reading or ±2 ppmv
- Repeatability: ±0.2 ppmv or ±0.1%, whichever is greater. For CO2 applications: ±1.0 ppmv or ±0.5%, whichever is greater
- Response Time: Optical system <2 seconds
- Operating Sample Cell Pressure: 10 to 25 psia (69 to 172 kPa)
Hazardous Area Certifications
- USA/Canada: Explosion-proof for Class I, Division 1, Groups B, C, D
- EU and Elsewhere: ATEX and IECEx: Ex de IIB+H2 T6 -20°C to +65°C Flameproof with increased safety compartment
A portable industrial TDLAS analyzer, our Panametrics Aurora TransPort delivers reliability and accuracy anywhere. This on-the-go moisture analyzer rapidly responds to changes in moisture content in most gases. The unit is assembled into a rugged and transportable case with a telescoping handle and wheels.
The TransPort is ideal for spot-checking the performance of natural gas processing and drying systems, gas storage facilities, compression stations, refinery processes, heat-treating furnaces, instrument air, and more. It is ideally suited for field verification of permanently installed moisture analyzers and transmitters. Features include:
Range
- Calibrated range: 0 to 5000 ppmv. For CO2 applications: 0 to 1000 ppmv
- Lower detection level: 2 ppmv (-71.7°C). For CO2 applications: 20 ppmv (-55.3°C)
- Process dew/frost point: Process/Equivalent dew Point/Frost point calculation at user-defined constant pressure or external pressure specified by Modbus
Accuracy
- Parts per million by volume: ±1% of reading or ±2 ppmv , whichever is greater; for >1000 ppmv ±5% of reading
- For CO2 applications: ±3% of reading or ±5 ppmv. For H2 recycle applications: ±1% of reading or ±2 ppmv (for up to ±5% H2 and ±1% C2H6 variation from nominal calibration composition) (Individual instrument calibrated accuracy conditions provided in Certificate of Conformance. Accuracy of other parameters derived from ppmv.)
- Repeatability: ±0.2 ppmv or ±0.1%, whichever is greater. For CO2 applications: ±1.0 ppmv or ±0.5%, whichever is greater
- Operating sample cell pressure 10 to 25 psia (69 to 172 kPa)
Battery Life
- Rechargeable battery provides 8 to 10 hours of operation
Panametrics Aurora RM rack-mount analyzer uses tunable diode laser absorption spectroscopy (TDLAS) to accurately measure moisture in a variety of background gases. The Aurora RM analyzer is suitable for installation in safe areas and operates over a wide range of environmental conditions.
Aurora RM’s fast response immediately alerts the user when moisture concentrations are out of compliance; once corrected, gas can be quickly cleared for process normalization.