- Ranges from 200 mbar to 700 bar
- Total accuracy to ±0.1 % FS
- Stainless steel construction
- Wide temperature range -40 °C to 125 °C
- Fast update rate 1 ms
- Customer set filter setting
- Baud rate upto 1000k bit/s
- Excellent long-term stability
Operating pressure ranges
Gauge ranges Any range from 200 mbar to 100 bar (3 to 1500 psi)
Ranges up 70 bar are enabled in negative gauge to the range or a maximum of -1 bar (-15 psi)
Absolute ranges Any range from 700 mbar to 700 bar (10 to 10000 psi) All ranges are zero based
Wet Dry differential Any range from 200 mbar to 35 bar (3 to 500 psi) All units are bidirectional Maximum line pressure: 70 bar
Wet Wet differential Any range from 700 mbar to 35 bar (10 to 500 psi) All units are bidirectional Maximum line pressure: 70 bar
Ordering When ordering, specify maximum working pressure. The lower pressure limit will be specified inline with rules stated above.
Over pressure At least 2 × Pressure range with negligible calibration change
For differential versions, the negative side must not exceed its positive side by more than: 2 × Pressure range up to a maximum of 15 bar
Containment pressure 4 × Pressure range for gauge ranges (200 bar (3000 psi) maximum) 200 bar (3000 psi) for absolute ranges below 100 bar (1500 psi) 1200 bar (17000 psi) for absolute ranges above 100 bar (1500 psi) For differential versions, the negative side must not exceed its positive side by more than: 2 × Pressure range up to a maximum of 15 bar
Supply Voltage: 5 Vdc to 32 Vdc Current: <30 mA
Output/communications CANopen V2.0B
CAN setting options (select when ordering) Node ID: Can be set to any integer between 2 and 127 (default setting is 2) Baud rate: Can be set to 10k, 20k, 50k, 125k, 250k, 500k, 800k, 1000k (default setting is 250k)
Power on time 500 ms to acquisition from power on
Insulation resistance >100 MΩ at 500 Vdc performance
Pressure performance Accuracy over the calibrated temperature range including zero and span setting and the effects of non-linearity, hysteresis and repeatability
Gauge/absolute/wet dry differential A3 Premium: ±0.1 % FS over -20 to 80 °C (-4 to 176 °F) A2 Improved: ±0.2 % FS over -40 to 125 °C (-40 to 257 °F)
Wet wet differential Premium: ±0.2 % FS over -20 to 80 °C (-4 to 176 °F) Improved: ±0.4 % FS over -40 to 125 °C (-40 to 257 °F) Increases pro-rata for pressures below 700 mbar
Temperature performance Accuracy over the calibrated temperature range ±3 °C (6 °F) with 0.1 °C (0.2 °F) Resolution
Long term stability ±0.05 % FS/year typical ±0.1 % FS maximum at reference conditions Increases pro-rata below 700 mbar (10 psi)
Line pressure effects (differential sensors only) Zero shift: <±0.03% span/bar Span shift: <±0.03% span/bar Effects increase pro-rata for ranges below 700 mbar
Environmental protection IP67
Operating temperature range -40 °C to 125 °C (-40 °F to 257 °F)
Pressure media Fluids compatible with stainless steel 316L and Hastelloy C276 Statement in accordance with the European pressure equipment directive
Enclosure materials Stainless steel 316L (body)
Pressure connectors
- PA G1/4 female
- PB G1/4 male flat
- PE 1/4 NPT female
- PF 1/4 NPT male
- PJ M14×1.5 60° Int Cone
- PZ M10×1 80° Int Cone
Other connectors may be available. Contact Druck to discuss your requirement.
Datasheets - DPS5000 CANBus